In Vivo Stability of Polyurethane-Based Electrospun Vascular Grafts in Terms of Chemistry and Mechanics Научная публикация
Журнал |
Polymers
ISSN: 2073-4360 |
||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|
Вых. Данные | Год: 2020, Том: 12, Номер: 4, Номер статьи : 845, Страниц : DOI: 10.3390/polym12040845 | ||||||||||
Ключевые слова | tecoflex; pellethane; gelatin; electrospinning; vascular grafts; polyurethane stability in vivo | ||||||||||
Авторы |
|
||||||||||
Организации |
|
Реферат:
The biostability of the polyurethanes Tecoflex EG-80A and Pellethane 2363-80A, used as basic polymers of the vascular grafts (VGs) produced by electrospinning, as well as the tensile strength of Tecoflex VGs, are studied. Solutions of Tecoflex or Pellethane with gelatin and bivalirudin in 1,1,1,3,3,3-hexafluoroisopropanol are used for VG production. After 1, 12, and 24 weeks of VG implantation in the infrarenal position of the abdominal aorta of Wistar rats, VGs are explanted, fixed in formalin, freed from outer tissues, dialyzed, and dried. The polyurethanes are extracted from VGs by dispersion/extraction in tetrahydrofuran (THF) and freed from the excess of THF-insoluble biopolymers. The stability of polyurethanes is assessed by IR spectroscopy and gel permeation chromatography. Pellethane has emerged to be stable at all experimental points. Tecoflex loses approximately 10% of its molecular weight (both M-n and M-w) after 3 months and restored its initial value within 6 months of its functioning as a graft. Mechanical testing demonstrates a 30% reduction in the tensile strength after 3 months in VG and a 10% increase after 6 months. The stability and mechanical properties of polyurethane-based VGs demonstrate their utility for the reconstitution of damaged arteries.
Библиографическая ссылка:
Gostev A.A.
, Shundrina I.K.
, Pastukhov V.I.
, Shutov A.V.
, Chernonosova V.S.
, Karpenko A.A.
, Laktionov P.P.
In Vivo Stability of Polyurethane-Based Electrospun Vascular Grafts in Terms of Chemistry and Mechanics
Polymers. 2020. V.12. N4. 845 . DOI: 10.3390/polym12040845 WOS Scopus РИНЦ
In Vivo Stability of Polyurethane-Based Electrospun Vascular Grafts in Terms of Chemistry and Mechanics
Polymers. 2020. V.12. N4. 845 . DOI: 10.3390/polym12040845 WOS Scopus РИНЦ
Файлы:
Полный текст от издателя
Даты:
Опубликована online: | 7 апр. 2020 г. |
Идентификаторы:
Web of science | WOS:000535587700115 |
Scopus | 2-s2.0-85086627614 |
РИНЦ | 43305984 |
OpenAlex | W3014554875 |